BIO-1100

Projection Perimeter

The Bio-1100 projection perimeter refers to Goldmann international standards, it integrates a variety of features. The background illumination, the size of the stimulus cursor and its brightness are standardized. Dynamic and static visual field detection can be performed, so that the whole picture of the visual field can be understood. It is used for the diagnosis and monitoring of glaucoma, vision and optic nerve diseases.

01

Impact body, Humanize design

Aspherical design, compact and humanize features, even when tested to 90 degrees, the patient's feeling is still very comfortable.

02

Equipped with a variety of fixation monitoring programs to ensure accurate measurement data and high reliability of measurement results

Built-in physiological blind spot detection, eye position video detection, eye position deviation alarm, and eye position deviation flexion line ,allow the operator to align and track the position of the patient's eyes and provide real-time assessment of eye gaze

03

Abundant inspection modes for both accuracy and speed

• Standards SWAP blue-yellow visual field examination,sensitive to earlier visual field defects
• Scientific and intelligent interactive inspection strategy, 30° range visual field inspection only takes 3 minutes

04

Comprehensive Analysis Software

• 7 kinds of analysis: reliability analysis, single-report analysis, multi-report analysis, glaucoma half-field analysis, glaucoma progression analysis, VFI visual field index analysis;blue-yellow visual field analysis
• Half-field analysis designed for glaucoma, making glaucoma diagnosis easier
• Analysis of glaucoma progression, very subtle progression is more visible and continues to indicate new progression

05

Internationally recognized six-figure report

Digital Map, Grayscale Map, Population Deviation, Population Deviation Probability Map, Mode Deviation, Mode Probability Map

Technical parameter

BIO-1100
Testing scope 90°
Testing distance 30cm
Background light white 31.5asb、yellow 315asb
Visual target 0.08asb~10000asb(0~51DB)
Target size Goldmann I, II, III, IV,V
Visual target stimulation time 200ms
Visual target interval time Standard and custom
Threshold policy Macula, 10-2, 24-2, 30-2, 60-4, nasal side
Upper threshold policy C-40,C-64,C-76,FF-81,FF-120,FF-135
Special detection strategy Esterman monocular, Esterman binocular, upper 36° .screening.blind point testing
Yellow-blue testing Standard 440nmblueV visual target+315asbOG 530 yellow background brightness
Visual target Red,blue
Dynamic policy Standard dynamic、blind dynamic、dark spot dynamic、straightline dynamic
Custom policy Static custom dynamic custom
Pupil size measurement Auto
Fixation monitoring Physiological blindspot monitoring, eye position.monitoring, eye position moving alarm, eye tracking curve
Environment light testing Auto
Analysis software Reliability analysis, single report, triple report, GHT glaucoma semi vision report, GPA glaucoma developing analysis, VFI analysis. Blue-yellow analysis
System Win 7 above
Others
Size 412mm(length) *497mm(width) *513mm(height)
Wight 17kg
Power 100-240VHZ:50-60Hz